System compensation is a required step in the VCSEL I-V Test
Subsystem setup process. You must
perform system compensation before you perform pulse compensation and I-V
measurements. After the initial setup is complete, you must perform system
compensation again if any connections within the VCSEL I-V Test
Subsystem change, such as a cable replacement, or if the facility
environment exceeds the environmental specifications for the subsystem.
Before you begin, complete the
following pre-requisites:
- Connect the PXI chassis to facility ground and a power source.
- Install an LO/Sense LO connector assembly on each PXIe-4135 in the PXI
chassis.
- Connect each NI-4134 to
the facility ground and a power source.
-
Connect each NI-4134 to a PXIe-4135 and the PXIe-8880.
-
Connect each NI-4134 to a PXIe-4135 and your computer.
NI recommends naming the BNC cables after the
NI-4134 pulse generator
outputs as follows and marking each BNC cable with an indicator so you can
differentiate between cables during compensation and general operation of the
VCSEL I-V Test
Subsystem.
- SCOPE cable—BNC cable that connects the load board to a 30 dB attenuator installed
on the PXIe-5162
front panel
- PRIMARY cable—BNC cable that connects the load board to the HIGH SPEED PULSE PRIMARY
output on the back panel of the NI-4134 pulse generator
- SECONDARY cable—BNC cable that connects the load board to the HIGH SPEED PULSE SECONDARY
output on the back panel of the NI-4134 pulse generator

Notice
Consistently connect the above BNC cables to the corresponding NI-4134 outputs during the
operation of the VCSEL I-V Test
Subsystem.
Otherwise, the compensation data and measurement results will be inaccurate.
Equipment
The following equipment is required to perform system
compensation on one subsystem channel.
- Male BNC short connector (x1)
- Female BNC short connector (x1)
- 30 dB attenuator (x1)
- BNC (m)-to-BNC (m) cables (x2 for full impedance NI-4134 configuration, x3
for half impedance NI-4134
configuration)
- PRIMARY cable
- SECONDARY cable (half impedance NI-4134
configuration only)
- SCOPE cable
Complete the following steps to perform system compensation.
-
Select from LabVIEW and navigate to to open VCSEL System and Pulse
Compensation.lvproj.
-
Open the System Compensation VI.
-
Follow the instructions on the VI front panel to complete system
compensation.
After the system compensation is complete, the
System Compensation VI automatically saves the system compensation
data for pulse compensation and I-V measurements.