Grounding the NI-4134

Last Modified: January 11, 2021

You must connect the NI-4134 grounding terminal to the grounding electrode system of the facility.

Required Equipment

  • Grounding wire
  • #2 Phillips screwdriver

Complete the following steps to ground the NI-4134.

  1. Remove the grounding screw from the grounding terminal on the NI-4134.
  2. Attach the ring lug to the grounding terminal.
  3. Tighten the grounding screw to 0.5 N⋅m (4.4 lb⋅in) of torque.
  4. Attach the other end of the wire to the system ground using a method that is appropriate for your application.
Repeat this procedure for each NI-4134 in the VCSEL I-V Test Subsystem.