Table Of Contents

Trigger Characteristics

Version:
    Last Modified: November 30, 2020

    Input triggers

    Types

    Start, Source, Sequence Advance, Measure, Pulse; Shutdown

    Sources (PXI trigger lines <0...7>)

    Polarity

    Configurable

    Minimum pulse width

    100 ns, nominal

    Destinations[2] (PXI trigger lines <0...7>)

    Polarity

    Active high (not configurable)

    Pulse width

    >200 ns, typical

    Output triggers (events)

    Types

    Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete, Pulse Complete, Ready for Pulse

    Destinations (PXI trigger lines <0...7>)

    Polarity

    Configurable

    Pulse width

    Configurable between 250 ns and 1.6 μs, nominal

    • 1 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.
    • 2 Input triggers can be re-exported.