Table Of Contents

Remote Sense

Version:
    Last Modified: November 30, 2020

    Voltage accuracy

    Add (3 ppm of voltage range + 11 µV) per volt of HI lead drop plus 1 µV per volt of lead drop per Ω of corresponding sense lead resistance to voltage accuracy specifications.

    Maximum sense lead resistance

    100 Ω

    Maximum lead drop per lead

    3 V, characteristic

    spd-note-note
    Note  

    Exceeding the maximum lead drop per lead value may result in additional error.