Table Of Contents

Output Resistance Programming Accuracy

Version:
    Last Modified: November 30, 2020

    Table 1. Output Resistance Programming Accuracy
    Current Level/Limit Range Programmable Resistance Range, Voltage Mode Programmable Resistance Range, Current Mode Accuracy ± (% of resistance setting), Tcal ± 5 °C [1]
    1 μA 0 to ±5 MΩ ±5 MΩ to ±infinity 0.03%
    10 μA 0 to ±500 kΩ ±500 kΩ to ±infinity
    100 μA 0 to ±50 kΩ ±50 kΩ to ±infinity
    1 mA 0 to ±5 kΩ ±5 kΩ to ±infinity
    10 mA 0 to ±500 Ω ±500 Ω to ±infinity
    100 mA 0 to ±50 Ω ±50 Ω to ±infinity
    1 A 0 to ±5 Ω ±5 Ω to ±infinity
    3 A 0 to ±500 mΩ ±500 mΩ to ±infinity
    10 A , pulsing only
    • 1 Tcal is the internal device temperature recorded by the PXIe-4139 at the completion of the last self-calibration.