Table Of Contents

Examples of Calculating Accuracy

Version:
    Last Modified: November 30, 2020

    spd-note-note
    Note  

    Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device.

    Example 1: Calculating 5 °C Accuracy

    Calculate the accuracy of 900 nA output in the 1 µA range under the following conditions:

    ambient temperature 28 °C
    internal device temperature within Tcal ± 5 °C []
    self-calibration within the last 24 hours.

    Solution

    Since the device internal temperature is within Tcal ± 5 °C and the ambient temperature is within 23 °C ± 5 °C, the appropriate accuracy specification is:

    0.03% + 100 pA

    Calculate the accuracy using the following equation:

    Accuracy = 900 nA * 0.03 % + 100 pA    = 270 pA + 100 pA

    = 370 pA

    Therefore, the actual output will be within 370 pA of 900 nA.

    Example 2: Calculating 1 °C Accuracy

    Calculate the accuracy of 900 nA output in the 1 µA range. Assume the same conditions as in Example 1, with the following differences:

    internal device temperature within Tcal ± 1 °C[]

    Solution

    Since the device internal temperature is within Tcal ± 1 °C and the ambient temperature is within 23 °C ± 5 °C, the appropriate accuracy specification is:

    0.022% + 40 pA

    Calculate the accuracy using the following equation:

    Accuracy = 900 nA * 0.022 % + 40 pA

    = 238 pA

    Therefore, the actual output will be within 238 pA of 900 nA.

    Example 3: Calculating Remote Sense Accuracy

    Calculate the remote sense accuracy of 500 mV output in the 600 mV range. Assume the same conditions as in Example 2, with the following differences:

    HI path lead drop 3 V
    HI sense lead resistance 2 Ω
    LO path lead drop 2.5 V
    LO sense lead resistance 1.5 Ω

    Solution

    Since the device internal temperature is within Tcal ± 1 °C and the ambient temperature is within 23 °C ± 5 °C, the appropriate accuracy specification is:

    0.016% + 30 μV

    Since the device is using remote sense, use the remote sense accuracy specification:

    Add (3 ppm of voltage range + 11 µV) per volt of HI lead drop plus 1 µV per volt of lead drop per Ω of corresponding sense lead resistance to voltage accuracy specifications.

    Calculate the remote sense accuracy using the following equation:

    Accuracy = ( 500 mV * 0.016 % + 30 μV ) + 600 mV * 3 ppm + 11 μV 1 V of lead drop * 3 V + 1 μ V V * Ω * 3 V * 2 Ω + 1 μ V V * Ω * 2.5 V * 1.5 Ω

       = 80 μV + 30 μV + 12.8 μV * 3 + 6 μV + 3.8 μV = 158.2 μV

    Therefore, the actual output will be within 158.2 µV of 500 mV.

    Example 4: Calculating Accuracy with Temperature Coefficient

    Calculate the accuracy of 900 nA output in the 1 µA range. Assume the same conditions as in Example 2, with the following differences:

    ambient temperature 15 °C

    Solution

    Since the device internal temperature is within Tcal ± 1 °C, the appropriate accuracy specification is:

    0.022% + 40 pA

    Since the ambient temperature falls outside of 23 °C ± 5 °C, use the following temperature coefficient per degree Celsius outside the 23 °C ± 5 °C range:

    0.0006% + 4 pA

    Calculate the accuracy using the following equation:

    Temperature Variation = ( 23 ° C 5 ° C ) 15 ° C = 3 ° C

    Accuracy = ( 900 nA * 0.022 % + 40 pA ) + 900 nA * 0.0006 % + 4 p A 1 ° C * 3 ° C

       = 238 pA + 28.2 pA

    = 266.2 pA

    Therefore, the actual output will be within 266.2 pA of 900 nA.

    • 1 Tcal is the internal device temperature recorded by the PXIe-4139 at the completion of the last self-calibration.

    Recently Viewed Topics