Table Of Contents

Output Resistance Programming Accuracy

Version:
    Last Modified: August 24, 2021

    Table 1. Output Resistance Programming Accuracy Characteristics
    Current Level/Limit Range Programmable Resistance Range, Voltage Mode Programmable Resistance Range, Current Mode Accuracy ± (% of resistance setting), Tcal ± 5 °C[1]
    1 μA 0 to ±5 MΩ ±5 MΩ to ±infinity 0.03%
    10 μA 0 to ±500 kΩ ±500 kΩ to ±infinity
    100 μA 0 to ±50 kΩ ±50 kΩ to ±infinity
    1 mA 0 to ±5 kΩ ±5 kΩ to ±infinity
    10 mA 0 to ±500 Ω ±500 Ω to ±infinity
    100 mA 0 to ±50 Ω ±50 Ω to ±infinity
    1 A 0 to ±5 Ω ±5 Ω to ±infinity
    3 A[2] 0 to ±500 mΩ ±500 mΩ to ±infinity
    • 1 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.
    • 2 3 A range above 1 A is for pulsing only.