Table Of Contents

Voltage

Version:
    Last Modified: August 16, 2021

    Table 1. Voltage Programming and Measurement Accuracy/Resolution
    Range Resolution (noise limited) Noise (0.1 Hz to 10 Hz, peak to peak), Typical Accuracy (23 °C ±5 °C) ± (% of voltage + offset)[1] Tempco ± (% of voltage + offset)/°C, 0 °C to 55 °C
    Tcal ±5 °C[2] Tcal ±1 °C[2]
    600 mV 100 nV 2 μV 0.020% + 50 μV 0.017% + 30 μV 0.0005% + 1 μV
    6 V 1 μV 6 μV 0.020% + 320 μV 0.017% + 90 μV
    20 V 10 μV 20 μV 0.022% + 1 mV 0.017% + 400 μV
    200 V 100 μV 200 μV 0.025% + 10 mV 0.020% + 2.5 mV
    • 1 Accuracy is specified for no load output configurations. Refer to Load Regulation and Remote Sense sections for additional accuracy derating and conditions.
    • 2 Tcal is the internal device temperature recorded by the PXIe-4135 at the completion of the last self-calibration.

    Recently Viewed Topics