This specifications apply to the PXIe-4113.
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
Specifications are valid under the following conditions unless otherwise noted.
Clean devices and terminal blocks by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with a soft, lint-free, dampened cloth. Do not use detergent or chemical solvents. The unit must be completely dry and free from contaminants before returning to service.
The PXIe-4113 is a single-quadrant power supply with two output channels that are capable of identical output. Channels are isolated from each other and from chassis ground.
|
|||||||
Minimum programmable voltage level/limit |
0.03 V |
||||||
|
|||||||
Output power |
60 W/channel |
The following figure illustrates the voltage and the current source capabilities of the PXIe-4113.
Specification | Resolution | Accuracy ± (% of output + offset), Tcal ± 5 °C[4] | Temperature Coefficient/°C, outside Tcal ± 5 °C |
---|---|---|---|
Voltage programming | 0.32 mV[2] | 0.1% + 15 mV | 0.004% + 0.1 mV |
Voltage measurement | 3 mV[3] | 0.1% + 15 mV | 0.004% + 0.25 mV |
Current programming | 0.19 mA[2] | 0.15% + 20 mA | 0.015% + 0.14 mA |
Current measurement | 2 mA[3] | 0.15% + 20 mA | 0.015% + 0.21 mA |
|
|||||||
|
Transient response |
<200 μs to recover within 100 mV of voltage level after a load current change from 50% to 100% of current range, typical |
||||||||||||||||||
|
|
Maximum output lead drop |
Up to 1 V drop per lead |
Maximum sense lead resistance |
Up to 1 Ω per lead |
|
|||||||
|
Voltage from auxiliary power + to earth ground |
60 VDC |
Voltage from auxiliary power - to earth ground |
1 VDC |
|
PXI power requirement |
0.5 A from 3.3 V rail, 0.25 A from 12 V rail |
Auxiliary power source input requirements |
45.6 VDC to 50.4 VDC, 3.5 A max |
|
|
|||||||||||||||||||||
|
Dimensions |
3U, one-slot, PXI Express/cPCI Express module; 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.), nominal |
Weight |
443 g, nominal |
User-replaceable fuse, auxiliary power input (PCB-mount) |
F 6.3 A H 250 V (5 × 20 mm ceramic fuse) |
|
Maximum altitude |
2,000 m (800 mbar) (at 25 °C ambient temperature) |
Pollution Degree |
2 |
Operating shock |
30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) |
||||||
|
The auxiliary power input port is sensitive to electrostatic discharge (ESD). When subjected to ESD during normal operation, a fault might result that requires user intervention to recover to normal operation. To ensure proper operation, make all I/O connections before attempting to use the device. In addition, take care to prevent ESD to the auxiliary power input port during normal operation.
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
For UL and other safety certifications, refer to the product label or the Online Product Certification section.
For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.
This product meets the essential requirements of applicable European Directives, as follows:
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.