These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
Specifications are valid under the following conditions unless otherwise noted.
Warranted specifications are valid under the following conditions unless otherwise noted.
Typical specifications are valid under the following conditions unless otherwise noted.
Number of channels |
Two (simultaneously sampled) |
Input type |
Referenced single-ended |
Connectors |
BNC, ground referenced |
Input impedance |
50 Ω ±1.5%, typical 1 MΩ ±1.0%, typical |
Input capacitance (1 MΩ) |
15.4 pF |
Input coupling |
AC DC |
Input Range (Vpk-pk) | Vertical Offset Range | |
---|---|---|
50 Ω | 1 MΩ | |
0.04 V | ±5 V | |
0.1 V | ±5 V | |
0.2 V | ±5 V | |
0.4 V | ±5 V | |
1 V | ±5 V | ±20 V |
2 V | ±5 V | ±20 V |
4 V | ±5 V | ±20 V |
10 V | ±2 V | ±100 V |
20 V | — | ±100 V |
40 V | — | ±100 V |
|
Resolution |
8 bits |
||||||||||||||||||||||
|
|||||||||||||||||||||||
DC drift[4] |
±[(0.2% × |Reading - Vertical Offset|) + (0.004% × |Vertical Offset|) + (0.013% of FS)] per °C |
||||||||||||||||||||||
AC amplitude accuracy[3] |
±0.25 dB at 50 kHz |
||||||||||||||||||||||
AC amplitude drift[4] |
±0.0026 dB per °C at 50 kHz |
|
|
Bandwidth-limiting filter |
20 MHz noise filter |
AC-coupling cutoff (-3 dB) |
10 Hz |
Supported triggers |
Reference (Stop) Trigger Reference (Arm) Trigger Start Trigger (Acquisition Arm) Advance Trigger |
||||||
Trigger types |
Edge Glitch Hysteresis Runt Width Window Digital Immediate Software |
||||||
Trigger sources |
CH 0 CH 1 PFI <0..3> PXI_Trig <0..7> |
||||||
|
|||||||
Trigger delay |
0 to 7.51 × 1014 ns [(251 - 1) * Sample Clock Period] |
||||||
Holdoff |
Dead time to 6.15 × 1018 ns [(264 - 1) * Sample Clock Period] |
Sources |
CH 0 CH 1 |
Interpolator Status | Time Resolution | |
---|---|---|
TIS Enabled | TIS Disabled | |
Enabled | 0.326 ps | 0.651 ps |
Disabled | 0.333 ns | 0.667 ns |
|
|||||||
Minimum threshold duration[17] |
Sample clock period |
Sources |
PFI <0..3> (front panel HD-BNC connectors) PXI_Trig <0..7> (backplane connector) |
||||||
|
Connectors |
PFI <0..3> (front panel HD-BNC connectors) |
||||||||||||||||
Direction |
Bidirectional per channel |
||||||||||||||||
|
|||||||||||||||||
|
Connectors |
Probe compensation terminal Ground terminal |
Output voltage[18] |
0 V to 5 V |
Maximum overload voltage |
25 V DC |
Available onboard memory sizes[21] |
64 MB 512 MB |
||||||
Minimum record length |
1 sample |
||||||
|
|||||||
Maximum number of records in onboard memory[22] |
100,000 |
Channels | Bytes per Sample | Maximum Records per Channel | Record Length | Allocated Onboard Memory per Record |
---|---|---|---|---|
1 | 1 | 100,000 | 1 | 192 |
1 | 1 | 100,000 | 1,000 | 1,200 |
1 | 1 | 52,758 | 10,000 | 10,176 |
1 | 1 | 1 | 536,870,784 | 536,870,976 |
2 | 1 | 100,000 | 1 | 192 |
2 | 1 | 100,000 | 1,000 | 2,208 |
2 | 1 | 26,630 | 10,000 | 20,160 |
2 | 1 | 1 | 268,435,392 | 536,870,976 |
External calibration corrects the onboard references for gain and offset errors used in self-calibration and adjusts the compensation attenuator. All calibration constants are stored in nonvolatile memory.
Self-calibration is done on software command. The calibration corrects for gain, offset, interleaving spurs, and intermodule synchronization errors. Run self-calibration after the specified warm-up time has elapsed and any time the module is power cycled or the PC or controller is restarted or wakes from sleep or hibernation modes. Refer to the NI High-Speed Digitizers Help at ni.com/manuals for more information on when to self-calibrate the device.
Driver support for this device was first available in NI-SCOPE 18.6.
NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXIe-5113. NI-SCOPE provides application programming interfaces for many development environments.
NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:
When you install NI-SCOPE on a 64-bit system, you can use InstrumentStudio to monitor, control, and record measurements from the PXIe-5113.
InstrumentStudio is an application that allows you to perform interactive measurements on several different NI device types in a single application.
Interactive control of the PXIe-5113 was first available via InstrumentStudio in NI-SCOPE 18.6. InstrumentStudio is included on the NI-SCOPE media.
NI Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5113. MAX is included on the driver media.
|
NI-TClk is an API that enables system synchronization of supported PXI modules in one or more PXI chassis, which you can use with the PXIe-5113 and NI-SCOPE.
NI-TClk uses a shared Reference Clock and triggers to align the Sample Clocks of PXI modules and synchronize the distribution and reception of triggers. These signals are routed through the PXI chassis backplane without external cable connections between PXI modules in the same chassis.
|
|||||||||||||||||||
Sample Clock delay/adjustment resolution |
<1 ps |
|
|||||||||
|
|||||||||
Total maximum power allowed |
30 W |
Dimensions |
3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in × 5.1 in × 8.5 in) |
Weight |
380 g (13.4 oz) |
|
|||||||
|
|||||||
Pollution Degree |
2 |
||||||
Maximum altitude |
4,600 m (at 25 °C ambient temperature) |
|
|||||||
Operating shock |
30 g, half-sine, 11 ms pulse |
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.