Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
Specifications in this document are Typical unless otherwise noted.
Specifications are valid under the following conditions unless otherwise noted.
Number of channels |
Two (simultaneously sampled) |
Connectors |
BNC |
|
|||||||
Input coupling |
Software-selectable: AC, DC, GND |
Range (Vpk-pk) | 50 Ω Offset (V) | 1 MΩ Offset (V) |
---|---|---|
0.1 | ±1 | ±1 |
0.2 | ||
0.4 | ||
1 | ||
2 | ±6 | ±10 |
4 | ±5 | |
10 | ±2 |
|
Resolution |
8 bits |
||||||||||||||||||
|
|||||||||||||||||||
Programmable vertical offset accuracy[2] |
±0.9% of offset setting, warranted |
||||||||||||||||||
|
|||||||||||||||||||
|
|
|||||||||||||||||||
|
|||||||||||||||||||
Bandwidth limit filter |
20 MHz noise filter |
||||||||||||||||||
|
|
|||||||
|
Range (Vpk-pk) | Noise Filter On | Noise Filter Off |
---|---|---|
0.1 | 240 µVrms (0.24% FS) | 320 µVrms (0.32% FS) |
0.2 | 480 µVrms (0.24% FS) | 600 µVrms (0.30% FS) |
0.4 | 960 µVrms (0.24% FS) | 1.12 mVrms (0.28% FS) |
1 | 2.4 mVrms (0.24% FS) | 2.6 mVrms (0.26% FS) |
2 | 4.8 mVrms (0.24% FS) | 6.0 mVrms (0.30% FS) |
4 | 9.6 mVrms (0.24% FS) | 11.2 mVrms (0.28% FS) |
10 | 24 mVrms (0.24% FS) | 26 mVrms (0.26% FS) |
Channel-to-channel skew |
<100 ps |
|
|
|||||||||
|
|||||||||
Sample clock delay range |
±1 Sample clock period |
||||||||
Sample clock delay/adjustment resolution |
≤5 ps |
Sources |
PFI 0 (front panel SMB connector) |
Frequency range[18] |
350 MHz to 1 GHz |
Duty cycle tolerance |
45% to 55% |
Sources |
PXI_CLK10 (PXI backplane connector) PFI 0 (front panel SMB connector) |
Frequency range[19] |
1 MHz to 20 MHz in 1 MHz increments Default: 10 MHz |
Duty cycle tolerance |
45% to 55% |
Exported Reference Clock destinations |
PXI_Trig <0..7> (backplane connector) PFI 1 (front panel SMB connector) |
Trigger types[20] |
Edge Window Hysteresis Video Digital Immediate Software |
||||||||
Trigger sources |
CH 0 CH 1 TRIG PFI <0..1> PXI_Trig <0..6> PXI Star Trigger Software |
||||||||
|
|||||||||
|
|||||||||
Holdoff |
From ream time up to [(232 - 1) × Sample clock period] |
||||||||
Trigger delay |
From 0 up to [(235 - 1) - Posttrigger samples] × (1/Sample rate), in seconds |
Trigger types |
Edge Window Hysteresis |
||||||
Sources |
CH 0 (front panel BNC connector) CH 1 (front panel BNC connector) TRIG (front panel BNC connector) |
||||||
|
|||||||
Voltage resolution |
8 bits (1 in 256) |
||||||
|
|||||||
|
|||||||
Trigger jitter[23] |
≤10 psrms, typical ≤20 psrms, maximum |
||||||
|
Trigger type |
Digital |
Sources |
PXI_Trig <0..6> (backplane connector) PFI <0..1> (front panel SMB connectors) PXI Star Trigger (backplane connector) |
Connector |
BNC |
Impedance |
1 MΩ in parallel with a nominal capacitance of 22 pF |
Coupling |
AC, DC |
AC coupling cutoff (-3 dB) |
12 Hz |
Input voltage range |
±5 V |
Maximum input overload |
|Peaks| ≤42 V |
Connector |
SMB jack |
Direction |
Bidirectional |
Destination |
Start trigger (acquisition arm) Reference (stop) trigger Arm reference trigger Advance trigger |
Input impedance |
150 kΩ, nominal |
VIH |
2.0 V |
VIL |
0.8 V |
Maximum input overload |
-0.5 V to 5.5 V |
Maximum frequency |
25 MHz |
Sources |
Start trigger (acquisition arm) Reference (stop) trigger End of record Done (end of acquisition) Probe compensation[24] |
Output impedance |
50 Ω |
Logic type |
3.3 V CMOS |
Maximum drive current |
±24 mA |
Maximum frequency |
25 MHz |
Real-Time and RIS Modes | Real-Time TIS Mode |
---|---|
8 MB standard (8 MS) per channel | 8 MB standard (8 MS) |
64 MB option (64 MS) per channel | 64 MB option (64 MS) |
256 MB option (256 MS) per channel | 256 MB option (256 MS) |
512 MB option (512 MS) per channel | 512 MB option (512 MS) |
Minimum record length |
1 sample |
||||||||||
Number of pretrigger samples |
Zero up to full record length |
||||||||||
Number of posttrigger samples |
Zero up to full record length |
||||||||||
|
|||||||||||
Allocated onboard memory per record |
[(Record length × 1 byte/sample) + 400 bytes] rounded up to next multiple of 128 bytes |
External calibration calibrates the VCSO and the voltage reference. All calibration constants are stored in nonvolatile memory.
Driver support for the PXI-5152 was first available in NI-SCOPE 3.2.
NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXI-5152. NI-SCOPE provides application programming interfaces for many development environments.
NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:
The NI-SCOPE Soft Front Panel (SFP) allows interactive control of the PXI-5152.
Interactive control of the PXI-5152 was first available in NI-SCOPE SFP version 3.2. The NI-SCOPE SFP is included on the NI-SCOPE media.
NI Measurement Automation Explorer (MAX) also provides interactive configuration and test tools for the PXI-5152. MAX is included on the NI-SCOPE media.
You can use the NI TClk synchronization method and the NI-TClk driver to align the Sample clocks on any number of supported devices, in one or more chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help. For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support.
Synchronization specifications are valid under the following conditions:
Skew[27] |
500 ps |
Skew after manual adjustment |
≤5 ps |
Sample clock delay/adjustment resolution |
≤5 ps |
Dimensions |
3U, one-slot, PXI module 21.6 cm × 2.0 cm × 13.0 cm (8.5 in × 0.8 in × 5.1 in) |
Weight |
462 g (16.3 oz) |
Maximum altitude |
2,000 m (at 25 °C ambient temperature) |
Pollution Degree |
2 |
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
For UL and other safety certifications, refer to the product label or the Online Product Certification section.
For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.
This product meets the essential requirements of applicable European Directives, as follows:
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.