Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
The following characteristic specifications describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
Specifications are valid at 23 °C unless otherwise noted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Topologies |
1-wire 4 x 128 matrix 1-wire 8 x 64 matrix 1-wire 16 x 32 matrix 1-wire dual 4 x 64 matrix 1-wire dual 8 x 32 matrix 1-wire dual 16 x 16 matrix 1-wire quad 4 x 32 matrix 1-wire sixteen 2 x 16 matrix 2-wire 4 x 64 matrix 2-wire 8 x 32 matrix 2-wire 16 x 16 matrix 2-wire dual 4 x 32 matrix |
This module is rated for Measurement Category I and intended to carry signal voltages no greater than 100 V. This module can withstand up to 500 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories II, III, or IV. Do not connect to MAINs supply circuits (for example, wall outlets) of 115 or 230 VAC.
|
|||||||||
Maximum current |
0.5 A (switching or carry, per channel) |
||||||||
Maximum switching power[1] |
10 W (per channel) |
||||||||
|
|||||||||
|
|||||||||
|
|||||||||
|
|||||||||
|
|
|||||||
|
Relay type[8] |
Reed |
||||||||||||||||||
Relay contact material |
Rhodium |
||||||||||||||||||
I/O connectors[9] |
2, 150 pos, Samtec ERM8-EM series |
||||||||||||||||||
|
|||||||||||||||||||
Dimensions (L × W × H) |
3U, one slot, PXI/cPCI module, PXI Express compatible 18.5 × 2.0 × 13.0 cm (7.3 × 0.8 × 5.1 in.) |
||||||||||||||||||
Weight |
454 g (1 lb) |
Operating temperature |
|
Storage temperature |
-20 °C to 70 °C |
Relative humidity |
5% to 85%, noncondensing |
Pollution Degree |
2 |
Maximum altitude |
2,000 m |
Indoor use only.
Operational Shock |
30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) |
||||||
|
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
For UL and other safety certifications, refer to the product label or the Online Product Certification section.
For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.
This product meets the essential requirements of applicable European Directives, as follows:
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.