Table Of Contents

Configure Spectrum Frequency (Start & Stop) (G Dataflow)

Version:
    Last Modified: December 19, 2018

    Configures the start and stop frequencies of a spectrum read by NI-RFSA.

    connector_pane_image
    datatype_icon

    session in

    Instrument session obtained from Initialize With Options .

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    start frequency

    Lower limit of a span of frequencies, in hertz (Hz).

    datatype_icon

    stop frequency

    Upper limit of a span of frequencies, in hertz (Hz).

    datatype_icon

    session out

    Reference to your instrument session passed to the next node.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Spectrum Span Behavior

    If you configure the spectrum span ( stop frequencystart frequency ) to a value larger than the instantaneous bandwidth of the device, NI-RFSA performs multiple acquisitions and combines them into a spectrum of the size you requested.

    PXIe-5663E Multispan Acquisition Behavior

    For the PXIe-5663E, NI-RFSA does not support multispan acquisitions from frequency ranges that correspond with different instantaneous bandwidths. For example, you cannot configure a multispan acquisition that acquires one span from 110 MHz to 120 MHz and a second from 120 MHz to 130 MHz because the bandwidths that correspond to each span are different (10 MHz and 20 MHz, respectively).

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported

    Web Server: Not supported in VIs that run in a web application

    Hardware Support:

    This node supports the following hardware:

    • PXIe-5601/5603/5605/5606 (external digitizer mode)
    • PXIe-5644/5645/5646
    • PXIe-5663E/5665/5667/5668
    • PXIe-5820/5840