Measures propagation delays through cables, connectors, and load boards using Time-Domain Reflectometry (TDR).
Ensure that the channels and pins you select are connected to an open circuit.
A Boolean that specifies whether to apply the measured TDR offsets.
True | TDR offsets are applied. |
False | TDR offsets are not applied. |
If you need to adjust the measured offsets prior to applying, set this input to False, and call the niDigital Apply TDR Offsets node to specify the adjusted TDR offsets values.
Default: True
Instrument session obtained from the niDigital Initialize with Options node.
List of channel names or list of pins. Do not pass a mix of channel names and pin names.
An empty string denotes all digital pattern instrument channels.
Specify channel names using the form PXI1Slot3/0,2-3 or PXI1Slot3/0,PXI1Slot3/2-3, where PXI1Slot3 is the instrument resource name and 0, 2, 3 are channel names.
To specify channels from multiple instruments, use the form PXI1Slot3/0,PXI1Slot3/2-3,PXI1Slot4/2-3. The instruments must be in the same chassis.
Site Considerations and Syntax
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Default: No error
Handle that identifies the session in all subsequent NI-Digital node calls.
Error information.
The node produces this output according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Where This Node Can Run:
Desktop OS: Windows
FPGA: Not supported
Web Server: Not supported in VIs that run in a web application