Adjusts the external calibration constants for a FD-11637 device. This node reports the as found error for gain and offset of the device.
calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.
physical channels specifies the physical channel(s) to calibrate.
reference value specifies the reference value measured using a calibrator.
error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.
actualReading indicates the value the device read during calibration. This value is used to determine gain and offset errors.
asFoundGainError indicates the calculated gain error.
asFoundOffsetError indicates the calculated offset error.
calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.
error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.