    Equipment

    • #2 Philips screwdriver
    • Adjustable wrench
    1. Power off the ECUTS-16000.
    2. Remove the ground cable nut from the PEP using the adjustable wrench.
    3. Remove the ground cable from the PEP.
    4. Disconnect the AC power cable, ethernet cable, and USB cable from the PEP.
    5. Remove the screws securing the USB ports to the PEP using the #2 Phillips screwdriver.
    6. Remove the screws securing the PEP to the test rack using the #2 Phillips screwdriver.
    7. Remove the PEP out from the test rack.
    8. Disconnect the ethernet cable, the Molex connector, and the power connector from the inside of the PEP.
    9. Connect the ethernet cable, the Molex connector, and the power connector to the inside of the replacement PEP.
    10. Place the PEP in the test rack.
    11. Screw in the screws securing the PEP to the test rack using the #2 Phillips screwdriver.
    12. Screw in the screws securing the USB ports to the PEP using the #2 Phillips screwdriver.
    13. Connect the AC power cable, ethernet cable, and USB cable from the PEP.
    14. Insert the ground cable into the PEP.
    15. Torque the ground cable nut to the PEP using the adjustable wrench to 1.29 N · m (11.5 lb · in.).