Replacing the SLSC Instrumentation

Version:
    Last Modified: March 29, 2021

    Equipment

    • #2 Phillips screwdriver
    • Flathead screwdriver
    1. Power off the ECUTS.
    2. Open the rear cabinet door.
    3. Loosen the cable screws on the front of the SLSC instrument using the flathead screwdriver.
    4. Disconnect all the cables from the SLSC instrument.
    5. Remove the two front panel mounting screws from the SLSC instrument using the #2 Phillips screwdriver.
    6. Eject the SLSC instrument from the chassis using the ejector handle at the bottom of the instrument.
    7. With the ejector handle engaged, slide the replacement SLSC instrument into the appropriate chassis slot.
    8. Lock the SLSC instrument in the chassis slot by pulling up on the ejector handle.
    9. Screw in the two front panel mounting screws on SLSC instrument using the #2 Phillips screwdriver.
    10. Connect all the cables to the front of the SLSC instrument.
    11. Screw in the cable screws on the front of the SLSC instrument using the flathead screwdriver.
    12. Close the rear cabinet door.