(ECUTS-16000) Replacing the PXI Chassis

Version:
    Last Modified: March 29, 2021

    Equipment

    • #2 Phillips screwdriver
    1. Power off the ECUTS-16000.
    2. Open the rear cabinet.
    3. Pull the spring pin attached to the inside hinge of the rear cabinet door down until it detaches from the test rack.
    4. Tilt and remove the rear cabinet door from the test rack.
    5. Remove the ground screw from the PXI chassis using the #2 Phillips screwdriver.
    6. Remove the rear PXI chassis feet brackets using the #2 Phillips screwdriver.
    7. While supporting the PXI chassis, carefully slide the PXI chassis out of the rear of the ECUTS-16000.
    8. While supporting the PXI chassis, carefully slide the replacement PXI chassis back into the rear of the ECUTS-16000.
    9. Replace the rear PXI chassis feet brackets using the #2 Phillips screwdriver.
    10. Replace the ground screw to the PXI chassis using the #2 Phillips screwdriver.
    11. Tilt and sit the rear cabinet door back in the test rack.
    12. Pull the spring pin attached to the inside hinge of the rear cabinet down and secure the pin back to the test rack.
    13. Close the rear cabinet door.