niRFSG 5611 Adjust Impairment Calibration VI
- Updated2025-10-14
- 3 minute(s) read
Calculates calibration information pertaining to the impairments of the device based on user-supplied measurements. The calibration information is stored in the driver session and can later be written to the device.
Supported Devices: PXIe-5611, PXIe-5673/5673E
Related Topics

Inputs/Outputs
max carrier leakage (dBc)
—
max sideband image (dBc)
—
instrument handle
—
instrument handle identifies your instrument calibration session. instrument handle is obtained from the niRFSG Initialize External Calibration VI.
measured RF OUT power (dBm)
—
measured RF OUT power specifies the power measured at the frequency indicated by the frequency to measure parameter in the niRFSG 5611 Configure Impairment Calibration VI.
error in (no error)
—
error in(no error) describes error conditions that occur before this node runs. The default is no error. This node does not pass the error in(no error) input to the error out output. Regardless of whether an error occurred before this node runs, the node returns no error. This input contains status, code, and source, which provide standard error in functionality.
instrument handle out
—
instrument handle out passes a reference to your instrument session to the next VI.
impairment calibration complete
—
impairment calibration complete indicates whether or not the impairment calibration is successful. If impairment calibration complete returns a value of TRUE, impairment calibration is successfully complete.
error out
—
error out contains error information. This output provides standard error out functionality. |
max carrier leakage (dBc)
—
instrument handle
—
error in (no error)
—
instrument handle out
—
impairment calibration complete
—
error out
—