NIRFSA_ATTR_SYNC_REF_TRIGGER_DIST_LINE
- Updated2025-05-22
- 1 minute(s) read
NIRFSA_ATTR_SYNC_REF_TRIGGER_DIST_LINE
Data type |
Access | High Level Functions |
---|---|---|
ViString | R/W | None |
Description
Specifies which external trigger line distributes the synchronized Reference Trigger signal. When synchronizing the Reference Trigger, configure all devices to use the same Reference Trigger distribution line.
Refer to the Synchronization Using NI-RFSA and NI-RFSG topic appropriate to your device in the NI RF Vector Signal Analyzers Help for more information about device synchronization for vector signal transceivers.
Valid Values: PXI_Trig0, PXI_Trig1, PXI_Trig2, PXI_Trig3, PXI_Trig4, PXI_Trig5, PXI_Trig6, PXI_Trig7, PFI0
Default Value: "" (empty string)
Supported Devices: PXIe-5644/5645/5646