RFSA C Reference

NIRFSA_ATTR_CAL_IF_FILTER_SELECTION

  • Updated2025-02-03
  • 1 minute(s) read

NIRFSA_ATTR_CAL_IF_FILTER_SELECTION

Specific Attribute

Data
type
Access High Level Functions
ViInt32 R/W None

Description

Specifies the IF filter path during calibration. The attribute is valid only during a calibration session.

Defined Values:

NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_1 (2100)    Specifies that the 5 MHz filter path is used during calibration.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_2 (2101) Specifies that the 300 kHz filter path is used during calibration. Not supported for the PXIe-5694.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_3 (2102) None of the IF filter paths are used during calibration.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_4 (2103) Specifies that the 20 MHz filter path is used during calibration.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_5 (2104) Specifies that the 1.4 MHz filter path is used during calibration.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_6 (2105) Specifies that the 400 kHz filter path is used during calibration.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_7 (2106) Specifies that the 110 kHz filter path is used during calibration.
NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_8 (2107) Specifies that the 30 kHz filter path is used during calibration.

Default Value: NIRFSA_VAL_EXT_CAL_IF_FILTER_PATH_4

Supported Devices: PXIe-5694