Indicates the minimum time between temperature sensor readings in seconds.

Syntax

NIRFSA_ATTR_TEMPERATURE_READ_INTERVAL

Numeric Value

Data Type

Access

Applies To

1150061

ViReal64

Read/Write

N/A

Remarks


When you call the niRFSA_ReadPowerSpectrumF64 function, the niRFSA_ReadIqSingleRecordComplexF64 function, or the niRFSA_Initiate function, NI-RFSA checks whether at least the amount of time specified by this attribute has elapsed before reading the hardware temperature.

Note

NI-RFSA ignores this attribute if you call the niRFSA_PerformThermalCorrection function or read the NIRFSA_ATTR_DOWNCONVERTER_GAIN attribute.

Default Value: 30 seconds

Supported Devices: PXI-5600, PXIe-5601/5603/5605/5606 (external digitizer mode), PXI-5661, PXIe-5663/5663E/5665/5667/5668, PXIe-5693/5694/5698, PXIe-5820/5830/5831/5832/5840/5841/5842/5860