RFmxNRMXAcpMeasurementMethod Enumeration
- Updated2025-04-08
- 2 minute(s) read
Specifies the method for performing the ACP measurement.
Syntax
Namespace: NationalInstruments.RFmx.NRMX
public enum RFmxNRMXAcpMeasurementMethod
Members
Name | Value | Description |
---|---|---|
Normal | 0 | The ACP measurement acquires the spectrum using the same signal analyzer setting across frequency bands. Use this method when measurement speed is desirable over higher dynamic range. |
DynamicRange | 1 | The ACP measurement acquires the spectrum using the hardware-specific optimizations for different frequency bands. Use this method to get the best dynamic range. Supported Devices: PXIe 5665/5668R. |
SequentialFft | 2 | The ACP measurement acquires I/Q samples for a duration specified by the ACP Sweep Time method. These samples are divided into smaller chunks. The size of each chunk is defined by the ACP Sequential FFT Size method, and the FFT is computed on each of these chunks. The resultant FFTs are averaged to get the spectrum and is used to compute the ACP. If the total acquired samples is not an integer multiple of the FFT size, the remaining samples at the end of the acquisition are not used for the measurement. Use this method to optimize ACP Measurement speed. The accuracy of results may be reduced when using this measurement method. For accurate power measurements when the power characteristics of the signal vary over time, averaging is allowed.
Note For multi-span FFT, the averaging count should be 1. |