RFmxNR_CfgRf
- Updated2025-04-08
- 1 minute(s) read
Configures the RF attributes of the signal specified by the selector string.
Syntax
int32 __stdcall RFmxNR_CfgRf(niRFmxInstrHandle instrumentHandle, char selectorString[], float64 centerFrequency, float64 referenceLevel, float64 externalAttenuation)
Parameters
Name | Direction | Type | Description |
---|---|---|---|
instrumentHandle | [in] | niRFmxInstrHandle | This parameter specifies the RFmx session refnum. Instrument Handle In is obtained from the RFmxInstr_Initialize function. |
selectorString | [in] | char | This parameter specifies a Selector String comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default is "" (empty string). |
centerFrequency | [in] | float64 | This parameter specifies the expected carrier frequency, in Hz, of the RF signal to acquire. The signal analyzer tunes to this frequency. The default of this attribute is hardware dependent. |
referenceLevel | [in] | float64 | This parameter specifies the reference level which represents the maximum expected power of the RF input signal. This value is expressed in dBm for RF devices and Vpk-pk for baseband devices. The default value of this parameter is hardware dependent. |
externalAttenuation | [in] | float64 | This parameter specifies the attenuation of a switch or cable connected to the RF IN connector of the signal analyzer. This value is expressed in dB. For more information about attenuation, refer to the RF Attenuation and Signal Levels topic for your device in the NI RF Vector Signal Analyzers Help. The default value is 0. |