Sets the temperature change required before RFmx recalculates the thermal correction settings when entering the running state. This value is expressed in degree Celsius.

Syntax

Namespace: NationalInstruments.RFmx.InstrMX

public int SetThermalCorrectionTemperatureResolution(string selectorString, double value)

Remarks

This method sets the value of ThermalCorrectionTemperatureResolution attribute.Supported devices: PXIe-5830/5831/5832/5840/5841/5842/5860 Default value: PXIe-5830/5831/5832/5842/5860: 0.2 PXIe-5840/5841: 1.0

Parameters

NameTypeDescription
selectorStringstring

Pass an empty string.

valuedouble

Specifies the temperature change required before RFmx recalculates the thermal correction settings when entering the running state.

Returns