Specifies the temperature change required before RFmx recalculates the thermal correction settings when entering the running state. This value is expressed in degree Celsius.

Syntax

RFMXINSTR_ATTR_THERMAL_CORRECTION_TEMPERATURE_RESOLUTION

Numeric Value

Data Type

Access

Applies To

120

float64

Read/Write

N/A

Remarks


You do not need to use a selector string if you want to configure this attribute for all signal instances. Specify the signal name in the selector string if you want to configure or read that signal instance. Refer to the Selector String topic for information about the string syntax.

Default value

PXIe-5830/5831/5832/5842/5860

0.2

PXIe-5840/5841

1.0

Supported devices: PXIe-5830/5831/5832/5840/5841/5842/5860