Configures channel power compensation for the inherent noise floor of the signal analyzer.


icon

Inputs/Outputs

  • cstr.png Selector String

    Selector String specifies a selector string comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default value is "" (empty string).

    Examples:

    ""

    "signal::sig1"

    You can use the RFmxCDMA2k Build Signal String VI to build the selector string.

  • cgenclassrntag.png Instrument Handle In

    Instrument Handle In specifies the RFmx session refnum. Instrument Handle In is obtained from the RFmxInstr Initialize NIRFSA VI.

  • ci32.png Noise Compensation Enabled

    Noise Compensation Enabled specifies whether to enable compensation of the channel powers for the inherent noise floor of the signal analyzer. The default value is False.

    False (0) Disables compensation of the channel powers for the noise floor of the signal analyzer.
    True (1) Enables compensation of the channel powers for the noise floor of the signal analyzer. The noise floor of the signal analyzer is measured for the RF path used by the ACP measurement and cached for future use. If the signal analyzer or measurement parameters change, noise floors are measured again.

    Supported Devices: PXIe-5663/5665/5668R, PXIe-5830/5831/5832/5842/5860

  • cerrcodeclst.png error in (no error)

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • igenclassrntag.png Instrument Handle Out

    Instrument Handle Out passes a reference to your RFmx session to the next VI.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.