RFmxCDMA2k ACP Configure Averaging VI
- Updated2025-04-08
- 3 minute(s) read
Configures averaging for the ACP measurement.

Inputs/Outputs
![]() Selector String specifies a selector string comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default value is "" (empty string). Examples: "" "signal::sig1" You can use the RFmxCDMA2k Build Signal String VI to build the selector string. ![]() Instrument Handle In specifies the RFmx session refnum. Instrument Handle In is obtained from the RFmxInstr Initialize NIRFSA VI. ![]() Averaging Enabled specifies whether to enable averaging for the measurement. The default value is False.
![]() Averaging Count specifies the number of acquisitions used for averaging when you set the Averaging Enabled parameter to True. The default value is 10. ![]() Averaging Type specifies the averaging type for averaging multiple spectrum acquisitions. RFmx uses the averaged spectrum for the measurement. The default value is RMS.
Refer to the Averaging section of the Spectrum topic for more information about averaging types. ![]() error in describes error conditions that occur before this node runs. This input provides standard error in functionality. ![]() Instrument Handle Out passes a reference to your RFmx session to the next VI. ![]() error out contains error information. This output provides standard error out functionality. |