RFmx VNA VIs

Table of Contents

RFmxVNA Calkit Manager Calkit Calibration Element Reflect Model Get C2 (VI)

  • Updated2024-04-10
  • 2 minute(s) read

RFmxVNA Calkit Manager Calkit Calibration Element Reflect Model Get C2 (VI)

Owning Palette: Reflect Model

Returns the 2nd order coefficient of the 3rd order polynomial capacitance/inductance model for the Reflect Open/Reflect Short model type.

RFmxVNA Calkit Manager Calkit Calibration Element Reflect Model Get C2

cio.gif

Instrument Handle In specifies the RFmx session refnum.

istr.gif

Selector String specifies the selector string created by this VI. The selector string comprises of the Calkit ID and the Calibration Element ID.

Example:

"ckit::MyCkitID/calel::MyCalelID"

You can use the RFmxVNA Build Calibration Element String VI to build the selector string.

ierrcodeclst.gif

error out returns error information.

cerrcodeclst.gif

error in describes error conditions that occur before this node runs.

iio.gif

Instrument Handle Out passes a reference of your RFmx session to the next VI.

idbl.gif

C2 (F or H) returns the 2nd order coefficient of the 3rd order polynomial capacitance/inductance model for the Reflect Open/Reflect Short model type.