Advanced:Limited Configuration Change Property
- Updated2023-10-12
- 4 minute(s) read
Advanced:Limited Configuration Change Property
Short Name: Limited Configuration Change
Property of RFmxTDSCDMA
Specifies the set of properties that are considered by RFmx in the locked signal configuration state.
If your test system performs the same measurement at different selected ports, multiple frequencies, and/or power levels repeatedly, you can enable this property to help achieve faster measurements. When you set this property to a value other than Disabled, the RFmx driver will use an optimized code path and skip some checks. Because RFmx skips some checks when you use this property, you need to be aware of the limitations of this feature, which are listed in the Limitations of the Limited Configuration Change Property topic.
You can also use this property to lock a specific instrument configuration for a signal so that every time that you initiate the signal, RFmx applies the RFmxInstr properties from a locked configuration.
NI recommends you use this property in conjunction with named signal configurations. Create named signal configurations for each measurement configuration in your test program and set this property to a value other than Disabled for one or more of the named signal configurations. This allows RFmx to precompute the acquisition settings for your measurement configurations and re-use the precomputed settings each time you initiate the measurement. You do not need to use this property if you create named signals for all the measurement configurations in your test program during test sequence initialization and do not change any RFmxInstr or personality properties while testing each device under test. RFmx automatically optimizes that use case.
Specify the named signal configuration you are setting this property in the selector string input. You do not need to use a selector string to configure or read this property for the default signal instance. Refer to the Selector String topic for information about the string syntax for named signals.
The default value is Disabled.
Disabled (0) |
This is the normal mode of RFmx operation. All configuration changes in RFmxInstr properties or in personality properties will be applied during RFmx Commit. |
No Change (1) |
Signal configuration is locked after the first Commit of the named signal configuration. Any configuration change thereafter either in RFmxInstr properties or personality properties will not be considered by subsequent RFmx Commits or Initiates of this signal. Use No Change if you have created named signal configurations for all measurement configurations but are setting some RFmxInstr properties. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode. |
Frequency (2) |
Signal configuration, other than center frequency and external attenuation, is locked after first Commit of the named signal configuration. Thereafter, only the Center Frequency and External Attenuation property value changes will be considered by subsequent driver Commits or Initiates of this signal. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode. |
Reference Level (3) |
Signal configuration, other than the reference level, is locked after first Commit of the named signal configuration. Thereafter only the Reference Level property value change will be considered by subsequent driver Commits or Initiates of this signal. If you have configured this signal to use an IQ Power Edge Trigger, NI recommends that you set the IQ Power Edge Level Type to Relative so that the trigger level is automatically adjusted as you adjust the reference level. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode. |
Freq and Ref Level (4) |
Signal configuration, other than center frequency, reference level, and external attenuation, is locked after first Commit of the named signal configuration. Thereafter only Center Frequency, Reference Level, and External Attenuation property value changes will be considered by subsequent driver Commits or Initiates of this signal. If you have configured this signal to use an IQ Power Edge Trigger, NI recommends you set the IQ Power Edge Level Type to Relative so that the trigger level is automatically adjusted as you adjust the reference level. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode. |
Selected Ports, Freq and Ref Level (5) |
Signal configuration, other than selected ports, center frequency, reference level, external attenuation, and RFInstr configuration, is locked after first Commit or Initiate of the named signal configuration. Thereafter only Selected Ports, Center Frequency, Reference Level, and External Attenuation property value changes will be considered by subsequent driver Commits or Initiates of this signal. If you have configured this signal to use an IQ Power Edge Trigger, NI recommends you set the IQ Power Edge Level Type to Relative so that the trigger level is automatically adjusted as you adjust the reference level. Refer to the Limitations of the Limited Configuration Change Property topic for more details about the limitations of using this mode. |
Remarks
The following table lists the characteristics of this property.
Datatype | ![]() |
Permissions | Read/Write |
High-level VIs | N/A |
Resettable | Yes |