RFMXTDSCDMA_ATTR_ACP_FAR_IF_OUTPUT_POWER_OFFSET

Data Type: float64
Access:read/write
Functions:RFmxTDSCDMA_SetAttributeF64
RFmxTDSCDMA_GetAttributeF64
Description: Specifies the offset by which to adjust the IF output power level for offset channels that are far from the carrier channel to improve the dynamic range. This value is expressed in dB. This attribute is used only if you set the RFMXTDSCDMA_ATTR_ACP_MEASUREMENT_METHOD attribute to RFMXTDSCDMA_VAL_ACP_MEASUREMENT_METHOD_DYNAMIC_RANGE and set the RFMXTDSCDMA_ATTR_ACP_IF_OUTPUT_POWER_OFFSET_AUTO attribute to RFMXTDSCDMA_VAL_ACP_IF_OUTPUT_POWER_OFFSET_AUTO_FALSE.

You do not need to use a selector string to configure or read this attribute for the default signal instance. Refer to the Selector Strings topic for information about the string syntax for named signals.

The default value is 20 dB.

Get Function: RFmxTDSCDMA_ACPGetFarIFOutputPowerOffset
Set Function: RFmxTDSCDMA_ACPSetFarIFOutputPowerOffset