RFmx LTE VIs

Table of Contents

RFmxLTE PvT Configure Averaging (VI)

  • Updated2024-04-10
  • 3 minute(s) read

RFmxLTE PvT Configure Averaging (VI)

Owning Palette: PVT

Configures averaging for the power versus time (PVT) measurement.

RFmxLTE PvT Configure Averaging

cio.gif

Instrument Handle In specifies the RFmx session refnum. Instrument Handle In is obtained from the RFmxInstr Initialize NIRFSA VI.

ci32.gif

Averaging Enabled specifies whether to enable averaging for the measurement. The default value is False.

False (0)

The measurement is performed on a single acquisition.

True (1)

The measurement is averaged over multiple acquisitions. The number of acquisitions is obtained by the Averaging Count parameter.

ci32.gif

Averaging Count specifies the number of acquisitions used for averaging when you set the Averaging Enabled parameter to True. The default value is 10.

ci32.gif

Averaging Type specifies the averaging type for averaging multiple spectrum acquisitions. The averaged spectrum is used for the measurement. The default value is RMS.

RMS (0)

The power spectrum is linearly averaged. RMS averaging reduces signal fluctuations but not the noise floor.

Log (1)

The power spectrum is averaged in a logarithmic scale.

cstr.gif

Selector String specifies a selector string comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default is "" (empty string).

Example:

"signal::sig1"

You can use the RFmxLTE Build Signal String VI to build the selector string.

cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

iio.gif

Instrument Handle Out passes a reference to your RFmx session to the next VI.

ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.

Log in to get a better experience