ACP:Noise Calibration:Averaging:Auto Property

  • Updated2024-04-10
  • 1 minute(s) read

ACP:Noise Calibration:Averaging:Auto Property

Short Name: ACP Noise Cal Averaging Auto

Property of RFmxLTE

Specifies whether RFmx automatically computes the averaging count used for instrument noise calibration.

You do not need to use a selector string to configure or read this property for the default signal instance. Refer to the Selector Strings topic for information about the string syntax for named signals.

The default value is True.

False (0)

RFmx uses the averages that you set for ACP Noise Cal Averaging Count property.

True (1)

RFmx uses the following averaging counts:

When you set the ACP Meas Method property to Normal or Sequential FFT, RFmx uses a noise calibration averaging count of 32.

When you set the ACP Meas Method property to Dynamic Range and sweep time is less than 5 ms, RFmx uses a noise calibration averaging count of 15.

When you set the ACP Meas Method property to Dynamic Range and sweep time is greater than or equal to 5 ms, RFmx uses a noise calibration averaging count of 5.

Remarks

The following table lists the characteristics of this property.

Datatype ci32.gif
PermissionsRead/Write
High-level VIs N/A
ResettableYes