RFmx for Bluetooth Test VIs

RFmxBT ModSpectrum Configure Averaging (VI)

  • Updated2024-04-10
  • 2 minute(s) read

RFmxBT ModSpectrum Configure Averaging (VI)

Owning Palette: ModSpectrum

Configures averaging for the ModSpectrum measurement.

RFmxBT ModSpectrum Configure Averaging

cio.gif

Instrument Handle In specifies the RFmx session refnum. Instrument Handle In is obtained from the RFmxInstr Initialize NIRFSA VI.

ci32.gif

Averaging Enabled specifies whether to enable averaging for the ModSpectrum measurement. The default value is False.

False (0) The measurement is performed on a single acquisition.
True (1) The measurement uses the Averaging Count parameter as the number of acquisitions over which the ModSpectrum measurement is averaged.
ci32.gif

Averaging Count specifies the number of acquisitions used for averaging when you set the Averaging Enabled parameter to True. The default value is 10.

cstr.gif

Selector String specifies a selector string comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default value is "" (empty string).

Examples:

""

"signal::sig1"

You can use the RFmxBT Build Signal String VI to build the selector string.

cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

iio.gif

Instrument Handle Out passes a reference to your RFmx session to the next VI.

ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.