Isolation Withstand Voltages
| Test section | Test voltage |
|---|---|
| 5 V communications power (logic), 24 V supply (I/O) | 500 VAC, 50 Hz, 1 min. |
| 5 V supply (logic) / analog inputs | 500 VAC, 50 Hz, 1 min. |
| 5 V supply (logic) / functional earth ground | 500 VAC, 50 Hz, 1 min. |
| 24 V supply (I/O) / analog inputs | 500 VAC, 50 Hz, 1 min. |
| 24 V supply (I/O) / functional earth ground | 500 VAC, 50 Hz, 1 min. |
| Analog inputs / functional earth ground | 500 VAC, 50 Hz, 1 min. |