De-embed Fixtures
- Updated2025-03-31
- 1 minute(s) read
De-embed the effects of all cabling and connectors between your DUT and the PXIe-5842 instrument to ensure that you are characterizing your DUT, not your larger test setup.
This process assumes your overall testing process uses an .s2p file to accomplish de-embedding.
Note If you are using the PXIe-5842 VST with
S-Parameters with a
connectorized fixture that can connect to the CAL-5501, you can perform measurement
calibration with the CAL-5501 so that the
measurement calibration plane is at the DUT. If this is the case,
de-embedding is not necessary.
This process describes de-embedding for 2-port devices.
You can de-embed elements of your physical setup with RFmxInstr Load S-parameter External Attenuation Table from S2P File.
Iterate through the following steps for each
port.
-
Specify, or allow an operator to input, the following
information:
- Name of the port in question
- Filepath to the .s2p file
- Orientation of the port with respect to the DUT: whether Port 1 or Port 2 of the .s2p is oriented towards the DUT
- Construct a port string with the port name entry and the Build Port String function in the RFmx API you are using.
- Call RFmxInstr Load S-parameter External Attenuation Table from S2P File.