Perform the following steps to self-calibrate the modules:
  1. Associate the PXIe-5841 and PXIe-5655 in MAX.
  2. Complete the steps described in the preceding Performing a Device Self-Calibration Using the NI-RFSA or NI-RFSG SFP section.
The composite self-calibration data of the PXIe-5841 with the PXIe-5655 does not replace the standalone PXIe-5841 calibration data. These two sets of calibration data can exist simultaneously. If you use the PXIe-5841 only, the PXIe-5841 uses its last unassociated self-calibration. If you use the PXIe-5841 associated with a PXIe-5655, the PXIe-5841 uses its last associated self-calibration.
Note When performing a standalone self-calibration, disconnect the PXIe-5655 from the PXIe-5841 or reset the PXIe-5655 in MAX to make sure the module is not generating.