Self-Calibration
- Updated2024-09-19
- 4 minute(s) read
Temperature drift and other environmental changes can lead to performance degradation for several PXIe-5840 specifications. Self-calibration adjusts the PXIe-5840 for variations in the module environment using an onboard high-precision calibration tone.
Perform self-calibration to compensate and optimize the performance of the PXIe-5840 for a given ambient temperature and operating environment.
- After first installing your PXIe-5840 into your chassis
- After any module in the chassis is added, moved, or removed
- When the system is in an environment where the ambient temperature varies
- If the module temperature has drifted more than ±5 °C from the last performed self-calibration
In addition to the preceding scenarios, performing proactive and regular self-calibration helps control the effects of aging-related drifts.
Self-calibration adjusts the following parameters of the PXIe-5840 to facilitate temperature correction:
- LO path gain
- RF input gain
- RF output gain
- RF input residual LO power
- RF output residual LO power
- RF input residual sideband image
- RF output residual sideband image
- RF input flatness
- RF output flatness
Self-Calibrating the PXIe-5840 with InstrumentStudio
NI recommends you perform an initial self-calibration using InstrumentStudio.
NI VSTs are represented in InstrumentStudio as a vector signal analyzer and a vector signal generator. Performing self-calibration with either representation of the PXIe-5840 accomplishes the same self-calibration, so it is not necessary to use both when you self-calibrate the PXIe-5840.
Complete the following steps to self-calibrate your VST with InstrumentStudio:
-
To ensure specifications before you use the instrument or perform self-calibration,
allow the instrument to warm up.
-
If it is not powered already, provide power to the instrument:
- : first provide power to the remote measurement module using the power adapter, and then power on the PXI Express chassis
- All other PXIe-5840 instruments: power on the PXI Express chassis
- Wait for the operating system to completely load.
- Allow the instrument to warm up for 30 minutes.
-
If it is not powered already, provide power to the instrument:
- Launch InstrumentStudio by navigating to .
- Click PXIe-5840. to create an analyzer panel for the
-
Click the tab created for the PXIe-5840 and click the
gear icon (
) in the upper right corner of the right sidebar.
- Hover over Self-Calibrate.... and click
- Click Continue to proceed.
Partial Self-Calibration
The partial self-calibration feature self-calibrates all configurations within the specified frequency and reference level limits of the PXIe-5840.
For best results, NI recommends that you perform partial self-calibration without omitting any steps. However, if certain aspects of performance are less important for your application, you can omit the following steps for faster execution:
- LO Self Cal—Omits the LO path gain adjustment.
- Amplitude Accuracy—Omits the RF absolute amplitude accuracy adjustment.
- Residual LO Power—Omits the RF residual LO power adjustment.
- Image Suppression—Omits the RF residual sideband suppression adjustment.
Performing Partial Self-Calibration Using LabVIEW
To perform a partial self-calibration when using LabVIEW, use the niRFSA Self Calibrate Range VI or the niRFSG Self Calibrate Range VI.
For more information about calibration, contact NI or visit ni.com/calibration. You can obtain the calibration certificate for your product at ni.com/calibration.
Related Information
- Device Warm-Up
NI recommends warming up the PXIe-5840 hardware for 30 minutes before operation.