Performing a Device Self-Calibration using the NI-RFSA SFP

NI recommends you perform the self-calibration from the NI-RFSA Soft Front Panel (SFP). You can also run a self-calibration programmatically using the NI-RFSA API by calling the niRFSA Self Cal VI or the niRFSA_SelfCalibrate function.

Complete the following steps to perform self-calibration using the NI-RFSA SFP.

  1. Launch the NI-RFSA SFP.
  2. Click Device/System » Calibration » Self Calibration.
  3. Select the self-calibration steps you want to perform from the Self-Calibration dialog box.
  4. Click Next to run the self-calibration.