Specifications are valid under the following conditions unless otherwise noted.

  • All vertical ranges
  • All bandwidths and bandwidth limit filters
  • Sample rate set to 1.25 GS/s, 2.5 GS/s, or 5 GS/s
  • Onboard Sample clock locked to onboard Reference clock

Warranted specifications are valid under the following conditions unless otherwise noted.

  • Ambient temperature ranges of 0 °C to 45 °C
  • The PXIe-5162 is warmed up for 15 minutes at ambient temperature
  • Self-calibration is completed after warm-up period
  • Calibration cycle is maintained
  • The PXI Express chassis fan speed is set to HIGH, the foam fan filters are removed if present, and the empty slots contain PXI chassis slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/docs.
  • NI-SCOPE 4.1 or later instrument driver is used
  • External calibration is performed at 23 °C ± 3 °C

Typical specifications are valid under the following conditions unless otherwise noted:

  • Ambient temperature ranges of 0 °C to 45 °C