Sample rate range

Real-time sampling (single shot)[1]1 Divide by n decimation used for all rates less than 100 MS/s.

1.526 kS/s to 100 MS/s

Random interleaved sampling (RIS)

200 MS/s to 2 GS/s in multiples of 100 MS/s

Phase noise density[2]2 10 MHz input signal.

<-100 dBc/Hz at 100 Hz

<-120 dBc/Hz at 1 kHz

<-130 dBc/Hz at 10 kHz

Sample clock jitter[3]3 Includes the effects of the converter aperture uncertainty and the clock circuitry jitter. Excludes trigger jitter.

1 psrms (100 Hz to 100 kHz)

2 psrms (100 Hz to 1 MHz)

Timebase frequency

100 MHz

Timebase accuracy

Not phase-locked to Reference clock

±25 ppm, Warranted

Phase-locked to Reference clock

Equal to the Reference clock accuracy

Sample clock delay range

±1 Sample clock period

Sample clock delay/adjustment resolution

≤10 ps

1 Divide by n decimation used for all rates less than 100 MS/s.

2 10 MHz input signal.

3 Includes the effects of the converter aperture uncertainty and the clock circuitry jitter. Excludes trigger jitter.