1. Restart the system.
  2. Launch MAX.
    • Failed self-test—Perform self-calibration, then perform the self-test again. The PXIe-4190 must be calibrated to pass the self-test.
    • Failed self-calibration—Perform self-calibration again.
  3. Power off the chassis.
  4. Reinstall the failed module in a different slot.
  5. Power on the chassis
  6. Perform the self-test again.