Accessory Specifications
- Updated2024-05-02
- 5 minute(s) read
Accessory Specifications
These specifications apply to the ACC-4162-1 or ACC-4163-1 while in use with the corresponding the PXIe-4162/PXIe-4163 SMU. These specifications supercede the relevant device specifications while the accessory is attached.
- The PXIe-4162 is a 12-Channel, ±24 V, 100 mA SMU.
- The PXIe-4163 is a 24-Channel, ±24 V, 50 mA SMU.
For complete device specifications refer to the PXIe-4162/PXIe-4163 specifications available at ni.com/docs.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Remote Sense
Maximum sense lead resistance | 100 Ω |
Maximum force lead drop | 1 V |
Nominal resistance between force and sense lines | 1 MΩ |
Voltage accuracy | Add 200 μV to the corresponding PXIe-4162/PXIe-4163 SMU accuracy specification when using a ACC-416x-1 accessory. |
Accessory Specifications
| |||||||
| |||||||
Load Regulation
For remote sense voltage accuracy (with the accessory attached), the lead drop is measured from the accessory output to the load.
For local sense voltage accuracy, consult the following table.
| Accessory | Typical Voltage | Maximum Voltage |
|---|---|---|
| ACC-4162-1 | 12 mV/mA | 17 mV/mA |
| ACC-4163-1 | 25 mV/mA | 32 mV/mA |
Physical
Dimensions | 75 mm × 81 mm × 19 mm (3.0 in. × 3.2 in. × 0.7 in.) |
Weight | 141 g (4.97 oz) |
Safety Voltage and Current
| |||||||
Environmental Guidelines
Environmental Characteristics
| |||||||||
| |||||||||
Pollution Degree | 2 | ||||||||
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) | ||||||||
| |||||||||
Compliance and Standards
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
Export Compliance
This product is subject to control under the U.S. Export Administration Regulations (15 CFR Part 730 et. seq.) administered by the U.S. Department of Commerce's Bureau of Industry and Security (BIS) (www.bis.doc.gov) and other applicable U.S. export control laws and sanctions regulations. This product may also be subject to additional license requirements of other countries' regulations.
Additionally, this product may also require export licensing before being returned to NI. The issuance of a Return Material Authorization (RMA) by NI does not constitute export authorization. The user must comply with all applicable export laws prior to exporting or re-exporting this product. See ni.com/legal/export-compliance for more information and to request relevant import classification codes (e.g. HTS), export classification codes (e.g. ECCN), and other import/export data.
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
1 Applies between HI and LO terminals. Leakage performance may be degraded for operation above 70% relative humidity. When transitioning a device from a storage or operation environment with operating humidity above 70%, allow the device to stabilize in the lower humidity environment for several hours before use.
2 Maximum resistance includes variation over operating temperature range.
3 When transitioning a device from a storage or operation environment with relative humidity above 70%, device should be allowed to stabilize in the lower humidity environment for several hours before use. Refer to the ACC-416x-1 accessory Voltage, Current, and Resistance specifications for additional performance derating when operating above 70% relative humidity.