Self-Calibrating the PXIe-4162 in MAX
- Updated2025-06-10
- 1 minute(s) read
Self-calibration adjusts the PXIe-4162 for variations in the module environment. The PXIe-4162 modules are externally calibrated at the factory, but you should perform a complete self-calibration after you install the module.
-
Install the PXIe-4162 and let it warm up for the
recommended warm-up time listed in the PXIe-4162
Specifications.
Note Warm up begins when the PXI chassis has been powered on and the operating system has completely loaded.
-
Self-calibrate the PXIe-4162 by clicking the
Self-Calibrate button in MAX or calling niDCPower Cal
Self Calibrate (niDCPower_CalSelfCalibrate).
Note Specify all channels of your PXIe-4162 with the channel name input when calling niDCPower Cal Self Calibrate (niDCPower_CalSelfCalibrate). You cannot self-calibrate a subset of PXIe-4162 channels.Note Low energy transients can appear at the output terminals of your PXIe-4162 during certain situations, such as power-up, power-down, device driver loading, and self-calibration.