S-Parameter Measurement Calibration Requirements
- Updated2025-05-13
- 1 minute(s) read
S-Parameter Measurement Calibration Requirements
Use the Calibration Wizard located inside the S-Parameter Interactive Example to perform system-level calibration of your S-parameter measurement test system.
Use the following table to determine which type of calibration is appropriate for your hardware configuration:
| Hardware Configuration | Calibration Type |
|---|---|
| 1 Coupler Per Port, 1 VST | SOLT |
| 1 Coupler Per Port, 2 VST | SOLT |
| 2 Couplers Per Port, 1 VST | QSOLT or SOLR |
| 2 Couplers Per Port, 1 VST (solid state switch) | QSOLT, SOLR, or TRL |
| S11, 1 Coupler | SOL |
| S11, 2 Couplers | SOL |
All types of calibration compensate for cabling losses, reflection at the VST RF ports and adapters, and the reflection and directivity of the couplers. NI recommends using the Maury Microwave 8050CK51 Characterized Device (CD) Fixed Load SOLT Kit with Adapters. The Pulsed RF Measurement Library provides a programming library and a wizard in the S-Parameter Interactive Example for this purpose.
Related Information
- Calibrating the System with the S-Parameter Interactive Example
Use the S-Parameter Measurements Interactive Example Calibration Wizard to guide you through a system-level calibration.