Pulsed RF Measurement Library API Reference

Table of Contents
  • S-Parameter Measurement VIs
  • PAE Measurement VIs
  • Pulse Measurement VIs

PRFMsc Configure RFmx Measurement Bandwidth

PRFMsc Configure RFmx Measurement Bandwidth

Configures the measurement bandwidth of the receiver. This VI sets the sampling rate as 1.25 times the value of Measurement Bandwidth (Hz). Sample rate should be an integer multiple of pulse repetition frequency (reciprocal of pulse repetition interval), so that the sampling locations of pulses are the same.

NoteNote: NI recommends setting the Measurement Bandwidth input to at least 10 divided by the pulse width.


Pulse Reference In

Specifies the session reference handle of the Pulse measurement session.

Measurement Bandwidth (Hz)

Specifies the expected measurement bandwidth of the received signal. The measurement bandwidth is used to set the IQ rate of the acquisition instrument. The default value is 2 MHz. The minimum setting should be 10 times the pulse width.

error in

Error conditions that occur before this node runs. This input provides standard error in functionality.

Pulse Reference Out

Returns the session reference handle of the Pulse measurement session.

error out

Error information. This output provides standard error out functionality.