Writes the reference level calibration data settings to the driver.

Supported Devices: PXIe-5601


icon

Inputs/Outputs

  • ci32.png ref level cal data type

    ref level cal data type specifies whether the reference level calibration data being used is the default configuration data or the mechanical relay disabled configuration data.

    Default (1800)

    The data is the default configuration data.

    Mechanical Attenuator Disabled (1801)

    The data is the configuration data when the mechanical relay is disabled. Use this option to save uncalibrated measurements for more advanced operation.

  • ci32.png RF band

    RF band specifies the RF band used during the reference level calibration.

    RF band 1 (1700)

    The RF band 1 path is used.

    RF band 2 (1701)

    The RF band 2 path is used.

    RF band 3 (1702)

    The RF band 3 path is used.

    RF band 4 (1703)

    The RF band 4 path is used.

  • civrn.png instrument handle

    instrument handle identifies your calibration instrument session. instrument handle is obtained from the niRFSA Init Ext Cal VI.

  • ci32.png attenuator table number

    attenuator table number specifies which attenuation table you are using. Valid values are 0 to 1.

  • cdbl.png frequency

    frequency specifies the frequency for the reference level adjustment.

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • cdbl.png measurement

    measurement specifies the relevant measurement taken for the current configuration.

  • iivrn.png instrument handle out

    instrument handle out passes a reference to your instrument session to the next VI. instrument handle is obtained from the niRFSA Initialize VI.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.