niRFSA Cal Adjust Device Gain VI
- Updated2025-10-14
- 2 minute(s) read
Records measured gain information that is gathered during the Reference Level Calibration step and IF Attenuation Calibration step.
This VI internally queries the properties you set, and you must commit all properties appropriate for your device calibration procedure prior to calling this VI. Refer to ni.com/manuals for the most recent version of the calibration procedure document for your device.
Call this VI immediately after a measurement is made and while the device under test (DUT) is still in the same state as it was during the measurement.
Supported Devices: PXIe-5693/5694/5698

Inputs/Outputs
instrument handle
—
instrument handle identifies your instrument session. instrument handle is obtained from the niRFSA Initialize VI.
frequency
—
frequency specifies the RF frequency, in Hz, of the measurement taken.
gain
—
gain specifies the gain measurement, in dB.
error in
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality.
instrument handle out
—
instrument handle out passes a reference to your instrument session to the next VI. instrument handle is obtained from the niRFSA Initialize VI.
error out
—
error out contains error information. This output provides standard error out functionality. |