Returns the miscellaneous information you can store during an external calibration using niScope Cal Store Misc Info.


icon

Inputs/Outputs

  • civrn.png instrument handle

    instrument handle is the instrument handle that you obtain from niScope Cal Start. The handle identifies a particular instrument session.

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

  • iivrn.png instrument handle out

    instrument handle out has the same value as the instrument handle.

  • istr.png miscellaneous information

    miscellaneous information is four characters that are stored in the EEPROM; however, it can be fewer than four characters if NULL-terminated.

  • ierrcodeclst.png error out