niRFSG 5653 Adjust LO Gain Calibration VI

  • Updated2024-05-08
  • 2 minute(s) read

niRFSG 5653 Adjust LO Gain Calibration VI

Calculates the PXIe-5653 LO gain calibration information. This calculation is based on user-supplied measurements. The calibration information is stored in the driver session, and you can later write this information to the device.

Supported Devices: PXIe-5653

niRFSG 5653 Adjust LO Gain Calibration

cio.gif

instrument handle identifies your instrument calibration session.

cdbl.gif

measured LO power specifies the power measured at the frequency indicated by the calibration frequency output of the niRFSG 5653 Configure LO Gain Calibration VI.

cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

iio.gif

instrument handle out passes a reference to your instrument session to the next VI.

ibool.gif

LO gain calibration complete indicates whether the LO gain calibration is complete.

ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.